Description and Specification for Lab Tenders
Water Purifier (RO/DI Water).
Water Output Type |
RO and DI water |
Water Output Speed |
10L/H |
Purifying Procedure |
PF+AC+RO+DI |
Water Supply Requirement |
Tap water: TDS<200ppm, 5~45℃, 1.0~3.5Kgf/cm2 |
Purification System |
Pre-treatment unit: 10’’PP filter*1+10’’Activated carbon *2 RO unit: 100GPD RO membrane*1 Subsequent unit: Deion purification column*2 |
Pure Water Quality |
●Soluble organic matter: Rejection rate>99% (molecular weight>100) ●Particle: Rejection rate>99%; ●Microbe: Rejection rate>99% ●Salt rejection rate:>99%; ●TDS (total solids solubility): RO water:5~10 ppm; ●Resistivity: 7.0~10.0 MΩ·cm; ●PM (particulate matter) (>0.22μm): <1/ml; ●Conductivity: 0.143~0.100 μs/cm; ●Microbe/Germ: <1 CFU/ml |
Water Quality Monitor |
TDS (Total Dissolved Solids) meter |
Consumption |
48W |
Power Supply |
AC110/220V±10%, 50/60Hz |
Standard Configuration |
Main body (Include 1 set of cartridges)+TDS meter |
Gross Weight |
38kg |
Categories: Water Purifier